Laser Particle Size Analyzer FM-LPA-A201

Description :

Laser Particle Size Analyzer FM-LPA-A201 is a dry dispersion laser particle size analyzer, adopted with MIE scattering theory, converging Fourier transform optical path, offers measuring range of 0.1 to 300 µm with 40 detector channels. Features He-Ne laser high-sensitivity photodetectors ring and semiconductor laser with high-precision feeding apparatus and Patent powder injection pumps. Equipped with intelligent operation system, dry turbulence dispersion and normal shock shearing effect, has unconstrained free fitting technology. Incorporated with different analysis mode, statistic method and comparison with <1 min of test time, provides test reports in various format and supports Win XP/Win 7 operation and USB interface.


Features:

    • A dry dispersion laser particle size analyzer adopted with MIE scattering theory
    • Converging Fourier transform optical path, ensures testing accuracy and reliability
    • Measuring range of 0.1 to 300 µm with 40 detector channels, reliable results
    • He-Ne laser high-sensitivity photodetectors ring and semiconductor laser
    • High-precision feeding apparatus and Patent powder injection pumps
    • Intelligent operation system, improve accuracy and authenticity of the testing results
    • Dry turbulence dispersion with normal shock shearing effect, sufficient dispersion
    • Unique unconstrained free fitting technology, accurate test results
    • Different analysis mode, statistic method and comparison with <1 min of test time
    • Test reports in various format, Win XP/Win 7 operation and USB interface
    • High-efficient, stable and reliable unit for stable, accurate and repeatable results

Specifications:

Dispersion Type Dry
Measurement Range 0.1 to 300 µm
Channel number 40
Accuracy Error <1% (National Standard Sample D50)
Repeatability Error <1% (National Standard Sample D50)
Light Source Semiconductor laser, λ = 632.8 nm
P > 2 mW
Detector High-sensitive ring photoelectric detector
Dispersion Mode Dry-turbulence dispersion mode
Dispersion Technique Normal shock wave shear technique
Feeding Automatic vibration feeding
Optical Path Alignment Fully automatic
Operation mode Automatic/ Manual (Switchable)
Standard ISO 13320-1:1999
Test Speed <1 min/time
Software Running WIN 7/10/XP, 64 bits
Analysis mode Free Distribution, R-R Distribution, Logarithm Normal Distribution, Mesh number classification etc.
Statistic Method Volume Distribution, Quantity Distribution
Test Report Word, Excel, Photo( Bmp), Text etc.
Interface USB
Dimension (L×W×H) 880×410×300 mm
Box 01 Packaging Dimension (L×W×H) 1090×610×500 mm
Box 02 Packaging Dimension (L×W×H) 410×390×570 mm
Box 03 Packaging Dimension (L×W×H) 830×450×770 mm
Net Weight 36 kg
Box 01 Gross Weight 62 kg
Box 02 Gross Weight 10 kg
Box 03 Gross Weight 60 kg

Applications:

Laser particle size analyzer is used to measure the sizes of particles in a material across cosmetics, food, coal, paint, paper, petrochemical, dyes, inks, explosives, kaolin, medicine, metal powder, mica, milling, minerals, oxides industries etc.


www.fison.com

Fison Instruments Ltd 272 Bath Street Glasgow G2 4JR UK