Double Beam Scanning UV-Visible Spectrophotometer FM-UVS-D100

Double Beam Scanning UV-Visible Spectrophotometer FM-UVS-D100

Double Beam Scanning UV-Visible Spectrophotometer FM-UVS-D100 is a compact, benchtop double beam optical system designed spectrophotometer comprising silicon photocell detector, offers 1.8 nm of spectrum bandwidth, and wavelength range of 190 to 1100 nm. Features Czerny-Turner monochromator with 160 mm focal length, tungsten and deuterium lamp as light source with sophisticated power protection system. Equipped with 8 inch color touch screen, USB COM port, and 1 position cell holder, this device offers Full-spectrum scanning, Linear regression, Detailed spectrum scanning, Concentration direct reading, Time-based kinetics determination, Peak / Valley detecting, GOTO λ, Multi-wavelength measurement. With automatic zero and full scale adjustment, automatic wavelength settings, Fast-Medium-Slow scanning speed, this device offers 1200 lines/mm grating.

  •  Compact, benchtop double beam optical system designed spectrophotometer
  •  Silicon photocell detector; Czerny-Turner monochromator
  •  Tungsten and deuterium lamp as light source with sophisticated power protection system
  •  8 inch color touch screen, USB COM port, and standard cell holder
  •  Offers Full-spectrum scanning, Linear regression, Detailed spectrum scanning, Concentration direct reading, Time-based kinetics determination, Peak / Valley detecting, GOTO λ, Multi-wavelength measurement
  •  Automatic zero and full-scale adjustment; Automatic wavelength settings
  •  Scanning speed: Fast-Medium-Slow

Wavelength range 190 to 1100 nm
Spectral bandwidth 1.8 nm
Optical system Double beam; Grating 1200 lines/mm
Wavelength accuracy ±0.5 nm
Wavelength repeatability ≤0.2 nm
Scanning speed Fast-Medium-Slow
Photometric range 0.0 to 200.0% T; -0.301 to 4.000 A; 0.000 to 9999C
Photometric accuracy ±0.3% T; ±0.002Abs (0 to 0.5 A); ±0.004 Abs (0.5 to 1 A)
Photometric repeatability ≤0.15% T; 0.001 Abs (0 to 0.5 A); 0.002 Abs (0.5 to 1 A)
Noise 100% (T) noise ≤0.15% (T) & 0% (T) noise≤0.1% (T)
Light source Tungsten lamp, Deuterium lamp
Stray light ≤0.03%T (at 220 nm NaI, 360 nm NaNo2)
Drifting ≤0.0009 Abs/30 min (250 nm and 500 nm after 2h warm up)
Display 8 inch color touch screen
Baseline ≤±0.002 A (200 to 1090 nm)
Standard Sample holder 1 position 10 mm cell holder
Sample compartment 10 mm pathlength cuvette
Detector Silicon Photocell
Monochromator Czerny-Turner
Focal length 160 mm
Function Full-spectrum scanning, Linear regression, Detailed spectrum scanning, Concentration direct reading, Time-based kinetics determination, Peak / Valley detecting, GOTO λ, Multi-wavelength measurement
Wavelength setting Automatic
Zero and full-scale adjustment Automatic
Interface USB
PC software Yes
Power supply AC 220 V,50 Hz, 200 W
Packaging Dimensions (W×D×H) 710×590×505 mm
Gross weight 36 kg

Double Beam Scanning UV-Visible Spectrophotometer are used for analysis of band gap, optical coatings and thin films, quantitative analyses, kinetics, wavelength scanning, and DNA and Protein analysis across clinical and research laboratories.

Glass cuvette 10 mm ×4
Quartz cuvette 10 mm ×2
Fuse ×2
Power cable