Potentiostat FM-PT-A100 ±10V potential control capability ensures high-precision electrochemical regulation. Dual 24-bit converters deliver exceptional signal clarity and operational stability. High-resolution sensing captures ultra-low current with advanced accuracy. Fast response architecture enhances dynamic test performance across ranges. Robust design guarantees consistent measurement reliability for demanding research.
| System Support | 2-, 3-, or 4-electrode system |
| Potential Control Range | ±10 V |
| Current Control Range | ±2 A |
| Potential Control Accuracy | 0.1%×full range ±1 mV |
| Current Control Accuracy | 0.1%×full range |
| Potential Resolution | 10 μV (>100 Hz), 3 μV (<10 Hz) |
| Current Sensitivity | 1 pA |
| Rise Time | <1 μs (<10 mA), <10 μs (<2 A) |
| Reference Electrode Input Impedance | 10¹² Ω |
| Current Range | 2 nA to 2 A, 10 ranges |
| Compliance Voltage | ±21 V |
| Maximum Current Output | 2 A |
| CV and LSV Scan Rate | 0.001 mV to 10,000 V/s |
| CA and CC Pulse Width | 0.0001 to 65,000 s |
| Current Increment During Scan | 1 mA at 1 A/ms |
| Potential Increment During Scan | 0.076 mV at 1 V/ms |
| SWV Frequency | 0.001 to 100 kHz |
| DPV and NPV Pulse Width | 0.0001 to 1000 s |
| AD Data Acquisition | 16-bit at 1 MHz, 20-bit at 1 kHz |
| DA Resolution | 16-bit, setup time: 1 μs |
| Minimum Potential Increment in CV | 0.075 mV |
| IMP Frequency | 10 μHz to 1 MHz |
| Low-Pass Filters | Covering 8-decade |
| Operating System | Windows 10/11 |
| Interface | USB 2.0 |
| Signal Generator - Frequency Range | 10 μHz to 1 MHz |
| Signal Generator - AC Amplitude | 1 mV to 2500 mV |
| Signal Generator - DC Bias | -10 to +10 V |
| Signal Generator - Output Impedance | 50 Ω |
| Signal Generator - Waveform | Sine wave, triangular wave, square wave |
| Signal Generator - Wave Distortion | <1% |
| Scanning Mode | Logarithmic/linear, increase/decrease |
| Signal Analyzer - Integral Time | Minimum: 10ms or cycle time, Max: 10⁶ cycles or 10⁵ s |
| Signal Analyzer - Measurement Delay | 0 to 10⁵ s |
| DC Offset Compensation - Potential Range | -10 V to +10 V |
| DC Offset Compensation - Current Range | -1 A to +1 A |
| Bandwidth | 8-decade frequency range, auto/manual setting |
| Product Dimension (L×B×H) | 360mm × 340mm × 140 mm |
| Packaging Dimension (L×B×H) | 550mm × 400mm × 250 mm |
| Net Weight | 6.5 kg |
| Packaging Weight | 9 kg |
Potentiostat FM-PT-A100 is ideal for corrosion science, electrocatalysis studies, battery research, materials analysis, electrochemical sensors, electrodeposition evaluation, energy-storage development, and precision measurement environments.
| Instrument host | 1 |
| CS studio software | 1 |
| Power cable | 1 |
| USB cable | 1 |
| electrode cable | 1 |
| Dummy cell | 1 |
| Manual | 1 |
| Ag/AgCl reference electrode | 1 |
| Glassy carbon working electrode (3mm) | 1 |
| Pt wire counter electrode | 1 |
| electrochemical cell 250 ml (non-sealed) | 1 |