Tap density tester FM-TDM-A100 is a single test station unit. It’s designed in accordance with USP I & USP II Tapped density Method. Measures tapped density, flowability of powder to optimize size of the packaging container. Tapped density is calculated in gm/ ml by dividing the sample weight by the final tapped volume.
Specifications:
Work station
1
Sample volume
≤ 250 ml
Sample weight
≤ 500 g
Number of Tapping
0 to 99999 (set 0 to output bulk density)
Frequency
300 rpm (continuous adjustable in 0 to 300 rpm)
Amplitude
Maximum 15 mm ( 1 to 15 mm integer variable)
Stroke height
Stroke height 3 mm
Strokes/ min
250 strokes/ min
Display
LED display
Repeatability
≤ 1 %
Accuracy
≤ 1 %
Humidity
≤ 85 % RH
Motor torque
0.86 N. m
Power
40 W
Power supply
AC 220 V± 10 %, 50 Hz
Features:
Test figure accuracy up to 4 decimal places
Stores test data in shut down mode
Highly advanced result interpretation
Built-in printer to print test report
Measures Hausner Ratio (TD/BD), Compressibility Index (TD-BD/TD × 100)
No need to adjust an integer
Automatically saves test result
Applications:
Used to test tapped density, simultaneously calculate bulk density, Hausner ratio across production, packaging, quality and research testing units in pharma, food, medicine, life science and other industries.
Complied Work Standards:
Sr. No.
Standards
1
ISO 3953: Metallic powder determination of tap density