Fison Automatic Bomb Calorimeter FM-ABC-B101 measures the heat released during sample combustion using a 300 mL oxygen bomb and temperature measurement system with 0.0001°C resolution. As an oxygen bomb calorimeter, it completes tests within ≤15 min and automates water filling, drainage, stirring, calorimetric measurement, and result printing.
| Standard | GB/T 213-2008 |
| Heat Capacity | 10,000 J/K |
| Oxygen Bomb Working Pressure | 2.8–3.0 MPa |
| Maximum Oxygen Bomb Pressure | 3.2 MPa |
| Oxygen Bomb Pressure Test | 20.0 MPa |
| Oxygen Bomb Volume | 300 mL |
| Oxygen Bomb Mass | 2.5 kg |
| Oxygen Bomb Dimensions | Ø86 × 181 mm |
| External Water Cylinder Capacity | 45 L |
| Inner Water Cylinder Capacity | 2100 mL |
| Ignition Voltage | AC 24 V |
| Ignition Mode | Fusible cotton wire ignition |
| Temperature Resolution | 0.0001°C |
| Temperature Measurement Range | 0–40°C |
| Measurement Accuracy | Meets GB/T 213-2008 |
| Test Time | ≤15 min |
| Power Supply | AC 220 V ±10%, 50 Hz |
| Power Consumption | <30 W in normal state; <300 W during ignition |
| Operating Environment | 5–35°C; humidity <80% |
| Display | Color touch screen |
| Operation | Automatic measurement process |
| Water Injection & Drainage | Automatic |
| Mixing | Automatic |
| Result Output | Automatic printing |
| Temperature Curve | Time-temperature curve displayed during testing |
| Water Injection Time Setting | 1–99 s |
| Drainage Time Setting | 1–99 s |
| Data Processing Units | cal/g, J/g, MJ/kg |
| Analytical Balance Requirement | Sensitivity 0.1 mg |
| Oxygen Requirement | ≥99.5% purity |
| Packaging dimensions | 76 x 57 x 48 cm |
| Net weight | 31kg |
| Gross weight | 35 kg |
Automatic Bomb Calorimeter FM-ABC-B101 used in coal, biomass, petroleum, chemical, food, wood, and material testing laboratories for determining the calorific value of combustible samples through controlled oxygen-bomb combustion.