Automatic Bomb Calorimeter FM-ABC-B101

Automatic Bomb Calorimeter FM-ABC-B101

Fison Automatic Bomb Calorimeter FM-ABC-B101 measures the heat released during sample combustion using a 300 mL oxygen bomb and temperature measurement system with 0.0001°C resolution. As an oxygen bomb calorimeter, it completes tests within ≤15 min and automates water filling, drainage, stirring, calorimetric measurement, and result printing.

$3,265.00


Download Catalog

Automatic Bomb Calorimeter FM-ABC-B101 - Specifications

Standard GB/T 213-2008
Heat Capacity 10,000 J/K
Oxygen Bomb Working Pressure 2.8–3.0 MPa
Maximum Oxygen Bomb Pressure 3.2 MPa
Oxygen Bomb Pressure Test 20.0 MPa
Oxygen Bomb Volume 300 mL
Oxygen Bomb Mass 2.5 kg
Oxygen Bomb Dimensions Ø86 × 181 mm
External Water Cylinder Capacity 45 L
Inner Water Cylinder Capacity 2100 mL
Ignition Voltage AC 24 V
Ignition Mode Fusible cotton wire ignition
Temperature Resolution 0.0001°C
Temperature Measurement Range 0–40°C
Measurement Accuracy Meets GB/T 213-2008
Test Time ≤15 min
Power Supply AC 220 V ±10%, 50 Hz
Power Consumption <30 W in normal state; <300 W during ignition
Operating Environment 5–35°C; humidity <80%
Display Color touch screen
Operation Automatic measurement process
Water Injection & Drainage Automatic
Mixing Automatic
Result Output Automatic printing
Temperature Curve Time-temperature curve displayed during testing
Water Injection Time Setting 1–99 s
Drainage Time Setting 1–99 s
Data Processing Units cal/g, J/g, MJ/kg
Analytical Balance Requirement Sensitivity 0.1 mg
Oxygen Requirement ≥99.5% purity
Packaging dimensions 76 x 57 x 48 cm
Net weight 31kg
Gross weight 35 kg

Automatic Bomb Calorimeter FM-ABC-B101 - Features

  • Intelligent MCU-based automatic calorimetric operation
  • Color touch screen display with user-friendly interface
  • Automatic control of the calorific value measurement process
  • Automatic water injection, mixing, ignition, drainage, and result printing
  • Automatic calorimetric system heat-capacity calibration
  • Time–temperature curve display during testing
  • Single-pump displacement water injection and drainage system
  • Electric stirring with flaky paddle design
  • Fusible cotton-thread ignition system
  • Built-in data processing for high and low calorific value calculations
  • Multiple calorific value units including J/g, MJ/kg, and cal/g
  • Automatic saving, displaying, and printing of test results
  • Dedicated system tests for ignition, mixing, water injection, and drainage
  • Designed according to GB/T 213-2008 requirements

Automatic Bomb Calorimeter FM-ABC-B101 - Applications

Automatic Bomb Calorimeter FM-ABC-B101 used in coal, biomass, petroleum, chemical, food, wood, and material testing laboratories for determining the calorific value of combustible samples through controlled oxygen-bomb combustion. 

Automatic Bomb Calorimeter FM-ABC-B101 - Standard Accessories

    • Oxygen bombs - 1
    • Oxygen filling instruments - 1
    • Oxygen decompressors, including oxygen delivery pipe - 1
    • Oxygen bomb racks - 1
    • Stainless steel crucible - 2, vulnerable spare parts
    • Vent valves - 1
    • 10 mL syringes - 1
    • Tweezers - 1
    • Scissors - 1
    • Scoop set - 1: large, medium, and small
    • Rubber sealing ring - 2, oxygen filling instrument, oxygen bomb one, Vulnerable spare parts
    • Thermal printing paper roll - 1: Consumables
    • Ignition wire roll 2: 0.18mm nickel chromium wire, consumables
    • Cotton thread - 1 Consumable
    • Benzoic acid tablet -10 Consumable