Gas chromatography FM-GC-A100 controls temperatures precisely from RT + 5°C to 450°C. It features a high-precision gas flow control system (EPC) for accuracy. It includes self-protection for gas circuit failure and automatic ignition. Innovative sampler design prevents sample sticking to the surface. Various injection methods like headspace sampling, thermal desorption analysis, and liquid automatic sampling are enabled. It can start analyses automatically at set times. The unit offers a user-friendly interface with a color widescreen and unique touch keys. Its modular design allows easy upgrades and compatibility with detectors such as FID, TCD, ECD, FPD, and NPD.
| Oven Temperature Range | RT + 2 to 450 °C |
| Temperature Control Accuracy | 0.01 °C |
| Temperature Program | ≥ 99 levels, 100 platforms |
| Maximum Heating Rate | 250 °C/min |
| Temperature Control System | 8-channel independent temperature control with automatic temperature protection function |
| Selectable Setector Types | FID, TCD, FPD, ECD, NPD (3 different types can be installed simultaneously) |
| Flame ionization detector (FID) | Operating temperature: 450 °C; detection limit: ≤ 1.1 pg C/s (n-hexadecane); baseline drift (30 min): ≤ 2 × 10⁻¹³ A; baseline noise: ≤ 2 × 10⁻¹⁴ A; dynamic linear range: ≥ 10⁷; automatic ignition function |
| Thermal Conductivity Detector (TCD) | Sensitivity: ≥ 10,000 mV·mL/mg; baseline drift (30 min): ≤ 0.2 mV; baseline noise: ≤ 0.05 mV; dynamic linear range: ≥ 10⁵; safety protection: gas bridging current protection function |
| Flame Photometric Detector (FPD) | Maximum operating temperature: 450 °C; limits of detection: P ≤ 80 fg/s, S ≤ 8 pg/s (methyl parathion); dynamic linear range: P ≥ 10³, S ≥ 10²; automatic ignition function |
| Electron Capture Detector (ECD) | Radioactive source: 63Ni; maximum operating temperature: 400 °C; limit of detection: ≤ 1.0 × 10⁻¹⁴ g/mL (γ-666); dynamic linear range: ≥ 10⁴ |
| Nitrogen-Phosphorus Detector (NPD) | Maximum operating temperature: 400 °C; limits of detection: N ≤ 1 × 10⁻¹² g(N)/s (azobenzene), P ≤ 5 × 10⁻¹³ g(P)/s (malathion); dynamic linear range: N ≥ 10³, P ≥ 10³ |
| Power Supply | AC 220 V, 50/60 Hz |
| Product Dimensions (W × D × H) | 670 × 580 × 500 mm |
| Packing Dimensions (W × D × H) | 840 × 720 × 750 mm |
| Net Weight | 60 kg |
| Gross Weight | 73 kg |
Gas chromatography is employed across diverse sectors for applications such as environmental monitoring, petrochemical analysis, pharmaceutical research, food and beverage testing, forensic investigations, chemical research, clinical diagnostics, material analysis, quality control, and research and development
FID Detector
Chromatography workstation software
Capillary Sampler
| Accessories Name | Specification |
|---|---|
| Sample Injection System | |
| Thermal Conductivity meter(TCD) | Sensitivity : > 5000 mv.ml /mg |
| ( 50 mg /ml toluene -benzene solution ) | |
| Temperature : 400°C | |
| Baseline drift : ≤ 0.1 m V /30 min | |
| Linear range : ≥ 105 ( ± 10 % ) | |
| Electron Capture Detector ( ECD ) | Detection limit : ≤ 1 × 10 -14 g / ml |
| ( Sample : n -hexadecane – δ 666 ) | |
| Linear Range : ≥ 10 4 | |
| Flame Photometric Detector ( FPD ) | Detection limit : ≤5 × 10 -12g/s (S of 10 ng/ul methyl parathion-ethanol solution) |
| Detection limit : ≤ 4×10 -12 (P of 10 ng/ul methyl parathion-ethanol solution) | |
| Linear range (S): ≥ 103 | |
| Linear range ( P) : ≥ 10 4 | |
| Nitrogen and Phosphorus Detector ( NPD ) | Detection limit : ≤ 1 × 10 -12g/s(Azobenzene) |
| Detection limit : ≤ 1 × 10-12 (Marla sulfur phosphorous) | |
| Linear range : ≥ 10 3 |