Fison Horizontal Autoclave FM-HA-B103 features a 1000 L stainless steel chamber with an operating temperature of up to 150°C and a maximum working pressure of 0.25 MPa for high-capacity steam sterilization. Equipped with automatic controls, an advanced vacuum system, integrated safety features, precise temperature regulation, and a durable pressure vessel, it supports efficient sterilization cycles with consistent performance.
| Capacity | 1000 L |
| Operating Temperature | 150 ℃ |
| Maximum Operating Temperature | 138 ℃ |
| Maximum Working Pressure | 0.25 MPa |
| Inner Chamber Design Pressure | -0.1 to 0.3 MPa |
| Safety Valve Opening Pressure (Chamber) | 0.28 MPa |
| Safety Valve Opening Pressure (Jacket) | 0.28 MPa |
| Sterilization Temperature Control Accuracy | ±0.2 ℃ |
| Temperature Uniformity | ±1 ℃ |
| Number of Pulsations | Amplitude -80 to 80 kPa, Frequency 1 |
| Positive Pressure Fluctuation Cycles | 1 to 3 |
| Vacuum Pulsation (Amplitude & Frequency) | Amplitude -80 to 0 kPa, Frequency 1 to 99 |
| Ultimate Vacuum | -97 kPa |
| Vacuum Leakage Rate | ≤0.13 kPa/min |
| External Surface Temperature | ≤ Ambient + 20 ℃ |
| Steam consumption | 55 Kg/C |
| Vessel Design Life | 10 years (20,000 sterilization cycles) |
| Internal Tank Material | 304-08 |
| Jacket Material | Q245R |
| Jacket Design Pressure | 0.3 MPa |
| Built-in Steam Generator (if any) | 0.3 MPa |
| Inner Chamber Hydrostatic Test Pressure | 0.39 MPa |
| Jacket Test Pressure | 0.52 MPa |
| Temperature Display Accuracy | 0.1 ℃ |
| Pressure Display Accuracy | 0.1 kPa |
| Dryness After Sterilization | Fabric ≤0.6%, Utensils/Tools ≤0.1% |
| Single Phase Voltage Supply | 220 ±10% V, 50 Hz |
| Three Phase Voltage Supply | 380 ±10% V, 50 Hz |
| Pure Steam (Self-produced) | 0.25 to 0.35 MPa (None) |
| Industrial Steam (Self-produced) | 0.3 to 0.5 MPa (None) |
| Water Softening | 0.15 to 0.3 MPa |
| Purified Water/Water for Injection | 0.15 to 0.3 MPa (for self-produced steam) |
| Compressed Air | 0.4 to 0.7 MPa |
| Depth | 1815 |
| Motor Power | 3.5 kW |
| Heating Power | 48 kW |
| Power Supply | 380 V, 50 Hz |
| Dimensions (L×W×H) | 2100 × 1350 × 2000 mm |
| Net Weight | 1850 kg |
| Gross Weight | 2150 kg |
| Packaging Dimensions | 239x152x228cm |
Horizontal Autoclave FM-HA-B103 is used for sterilizing laboratory instruments, clothing, rubber stoppers, filters, culture media, and other reusable materials. It is suitable for hospitals, pharmaceutical and biotechnology facilities, and research laboratories.