Class II Biosafety Cabinet FM-BSC-A407 ensures high-efficiency air purification with ULPA filtration and 480 mm maximum opening. It features motorized front window to enable smooth and hands-free operation for enhanced convenience. It is designed with the advanced interlocking system to ensure user protection and prevent contamination. Its digital LCD screen provides real-time safety alerts and system status updates. Our Biosafety cabinet offers a spacious workspace with customizable base height.
Specifications
Max opening | 480 mm |
Tested opening | Safety height 200 mm (8”) |
Inflow velocity | 0.53±0.025 m/s |
Downflow velocity | 0.33±0.025 m/s |
ULPA filter | Two, 99.9995 efficiency at 0.12µm, filter life indicator |
Front window | Motorized, two-layer laminated toughened glass ≥6 mm, anti UV |
Noise | ≤65dB |
UV lamp | 30W*1 UV timer, UV life indicator, emission of 253.7 nanometres for most efficient decontamination |
LED lamp | 14W*2 |
Illumination | ≥1000Lux |
Waterproof socket | Two, total load of two sockets: 500W |
display | LCD display: exhaust filter and down flow filter pressure, filter and UV lamp filter time, inflow and downflow velocity, filter life, humidity and temperature, system working time etc. |
Control system | Microprocessor |
Airflow system | 70% air circulation, 30% air exhaust |
Visual and audio alarm | Abnormal flow velocity, filter replacement, front down at unsafe height, high filter pressure alarm, abnormal power failure |
Material | Work zone: 304 stainless steel Main body: cold rolled steel with anti-bacteria powder coating |
Work surface height | 770 mm (can be customized) |
Caster | Footmaster caster |
Power consumption | 800W |
Power supply | AC220V, 50/60Hz; 110V, 60Hz |
Internal dimension | 1150×600×660 mm |
External dimension | 1300×755×2200 mm |
Packaging dimension | 1430×990×1840 mm |
Net weight | 260 kg |
Gross weight | 280 kg |
Features
Applications
Accessories
LED lamp × 2
UV lamp × 2
Base stand × 1
Remote control × 1
Foot switch × 1
Drain valve × 1
waterproof socket × 2
Accessories Optional
Water and gas tap
Electric height adjustable base stand
Armrest
Fison Class II Biosafety Cabinets offer personnel, product, and environmental protection by preventing contamination during biological sample handling. They are equipped with a laminar airflow design that maintains continuous airflow for optimal sample protection. Our cabinets feature a HEPA filtration system that effectively removes airborne contaminants. They provide negative pressure to prevent hazardous particle escape. These units deliver energy-efficient operation with optimized airflow and lighting for safety and performance.