Fison Atomic Absorption Spectrophotometers offer high precision and sensitivity for detecting metal concentrations in various samples. They feature an advanced optical system with a high-resolution monochromator for enhanced spectral accuracy. These units provide automated sample handling, reducing manual intervention. They include a user-friendly interface and digital controls for easy operation. Our spectrophotometers deliver multiple calibration modes, allowing single and multi-element analysis to ensure versatility.
Main specification
Wavelength range
Light source system
Wavelength accuracy
Optical system
Wavelength range | 190 to 900 nm |
Wavelength accuracy | ±0.5 nm |
Resolution | Two spectral lines of Mn at 279.5 nm and 279.8 nm, can be separated with spectral bandwidth of 0.2 nm and valley-peak energy ratio less than 30% |
Baseline stability | 0.005 A/30 min |
Background correction | D2 lamp background correction capability at 1A is better than 30 times |
Lamp turret | 2 lamps powered simultaneously (one preheated) |
Lamp current adjustment | 0 to 20 mA |
Lamp power supply mode | 400 Hz square wave pulse |
Monochromator | Single beam, Czerny-Turner design grating monochromator |
Grating | 1800 lines/mm |
Focal length | 277 mm |
Blazed wavelength | 250 nm |
Spectral bandwidth | 0.1 nm, 0.2 nm, 0.4 nm, 1.2 nm, automatic change |
Spectral adjustment | Manual adjustment for wavelength and slit |
Burner | 10 cm single slot all-titanium burner |
Spray chamber | Corrosion resistant all-plastic spray chamber |
Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate:6-7 ml/min |
Position adjustment | Manual adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner/min |
Gas line protection | Fuel gas leakage alarm/min |
Wavelength range | 190 to 900 nm |
Wavelength accuracy | ± 0.25 nm |
Resolution | Two spectral lines of Mn at 279.5 nm and 279.8 nm can be separated with the spectral bandwidth of 0.2 nm and valley peak energy ratio less than 30% |
Baseline stability | ≤0.004 A/30 min |
Background correction | D2 lamp background correction capability at 1A is better than 30 times |
Lamp turret | 4-lamp turret auto alignment, fully automated scan and peak-picking |
Lamp current adjustment | Automatic adjustment and display Wide pulse current: 0 to 25 mA, Narrow pulse current: 0 to 10 mA |
Lamp power supply mode | 400 Hz square wave pulse |
Monochromator | Single beam, Czerny-Turner design grating monochromator |
Grating | 1800 lines/mm |
Focal length | 277 mm |
Blazed wavelength | 250 nm |
Spectral bandwidth | 0.1 nm, 0.2 nm, 0.4 nm, 1.2 nm automatic change |
Burner | 10 cm single slot all-titanium burner |
Spray chamber | Corrosion resistant all-plastic spray chamber |
Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate: 6 to 7 mL/min |