Fluorometer FM-FLM-A103 handles a sample volume range of 1 to 20 µL, providing reliable fluorescence detection. It operates at excitation wavelengths of 470 nm and 625 nm, ensuring accurate detection of fluorescence signals across a broad spectral range. It is equipped with two detection channels and supports single-sample analysis, ensuring accurate results. It uses 0.5 ml PCR tubes, providing compatibility for easy and efficient sample analysis. It features a touch screen display for real-time monitoring of key parameters.
| Sensitivity | 0.5 pg/µL dsDNA |
| Sample Capacity | 1 |
| Excitation Wavelength | 470/625 nm (standard), 365/525 nm (optional) |
| Emission Wavelength | 525/690 nm (standard), 460/620 nm (optional) |
| Concentration Range | dsDNA HS: 0.005 ng/µL to 120 ng/µLdsDNA BR: 0.2 ng/µL to 2000 ng/µLOligo: 0.01 ng/µL to 240 ng/µLRNA: 0.2 ng/µL to 200 ng/µLMicroRNA: 0.025 ng/µL to 150 ng/µLProtein: 12.5 µg/ml to 5 mg/mlProtein BR: 0.1 mg/ml to 20 mg/ml |
| Light Source | Monochrome LED |
| Sample Volume | 1 to 20 µL |
| Detector | Photodiode |
| Tube Type | 0.5 ml PCR tube |
| Channel Number | 2 |
| Programs Stored | 10,000 |
| Measurement Time | 3 seconds |
| Data Export | U Disk |
| Repeatability | < 1.5% |
| Data Format | CSV, PDF |
| Calibration Method | 2 or 3-point calibration |
| Data Interface | USB (Type A) x 2, USB (Type B) x 1 |
| Response Range | Five orders of magnitude |
| Power Adapter | 100 to 240 V, 50 to 60 Hz |
| Linearity | R² ≥ 0.995 |
| Power Consumption | 4.5 W |
| Power Supply | DC 12 V, 2 A |
| Weight | 1.0 kg |
Fluorometer FM-FLM-A103 is used in oceanography to measure chlorophyll concentrations. It is widely used in life sciences, environmental monitoring, pharmaceutical quality control, and bioengineering applications.