Fison FTIR Spectrometer FM-FTI-A102 offers a spectral range of 7800 to 350 cm⁻¹ with high SNR up to 20,000:1 and resolution of 0.85 cm⁻¹, ensuring accurate infrared analysis. It features DLATGS detector, cube-corner Michelson interferometer, Ethernet connectivity, and advanced software for data processing, report generation, and real-time monitoring, making it suitable for laboratory, research, and quality control applications.
$30,437.00
| Interferometer | Cube-corner Michelson interferometer |
| Beam Splitter | Multilayer Ge coated KBr |
| Detector | Room temperature DLATGS module (standard) |
| IR Source | High intensity, long lifetime, air-cooled IR source |
| Wavenumber Range | 7800 cm-1 to 350 cm-1 |
| Resolution | 0.85 cm-1 |
| Signal To Noise Ratio | Better than 20,000:1 (RMS value, at 2100 cm-1 to 2000 cm-1 or 2100 cm-1 to 2200 cm-1, Resolution: 4 cm-1, Detector: DLATGS, 1 minute data collection |
| Wavenumber Accuracy | ± 0.01 cm-1 |
| Scanning Speed | Microprocessor control, different scanning speed selectable |
| Software | mainFTOS Suite software workstation, compatible to all version Window OS |
| Communication | Ethernet interface (standard) |
| Data Output | Standard data format, report generation and output |
| Status Diagnosis | Power on self-check, real-time temperature and humidity monitoring and reminders |
| Environment Condition | Temperature: 10°C to 30°C, Humidity: less than 60 % |
| Power Supply | AC 220 V ± 22 V, 50 Hz ± 1 Hz |
| Dimensions | 490×420×240 mm |
| Weight | 24 kg |
FTIR Spectrometer FM-FTI-A102 is used for material identification and compositional analysis in pharmaceuticals, biopharma, metals and mining, geochemistry, energy and fuels, and forensic and drug testing. It is also widely used in industrial quality control, research laboratories, and production testing for reliable and accurate spectral analysis.
Transmission sample holder
| Accessory | Quantity |
| KBr powder analytical grade | 500g/bottle |
| KBr powder spectroscopic grade | 25g/bottle |
| KBr powder spectroscopic grade | 10g/bottle |
| Diffuse/Specular reflectance | set |
| Horizontal ATR (ZnSe or KRS-5 crystal) | set |
| Fixed sample holder - 10mm Cuvette | PC |
| Demoutable sample holder - 10-40mm Cuvette | PC |
| Diffuse transmission accessory | 1 set |
| Hydroxyl measurement accessory (automatic analysis system for measuring hydroxyl content in the quartz glass) | 1 set |
| Liquid ATR(Import ZnSe crystal) | 1 set |
| Single reflection ATR (ZnSe crystal) | 1 set |
| Single reflection ATR (Diamond crystal) | 1 set |
| Mortar and pestle(Φ70) | 1 PC |
| Mortar and pestle(Φ50) | 1 PC |
| Liquid cells(KBr) | 1 PC |
| Liquid cells(CaF2) | 1 PC |
| Gas cells | 1 PC |
| KBr window plate(Φ25ⅹ4) | 1 PC |
| CaF2 window plate(Φ25ⅹ4) | 1 PC |
| BaF2 window plate(Φ25ⅹ4) | 1 PC |
| ZnSe window plate(Φ25ⅹ2) | 1 PC |
| KBr window plate polishing kit | 1 PC |
| Computer - Lenovo dual-core 2GRAM+500Gharddisk+DVD +19 "LCD | 1 PC |
| Printer - HP-1108 | 1 PC |
| Hydraulic press 15T (including die set) | 1 PC |
| Dehumidifier -(50L) | 1 PC |
| Die set - 15T | 1 PC |
| Infrared oven | 1 PC |
| Temperature controller and heater | 1 PC |
| Intelligent electronic moisture-proof box | 1 Pc |
| Sr. No | Advance Tools Name | Description | |
| 1. | Internet + Detection instrument design with multiple communication options: | I. | Adapting to “Internet+ detection” instrument development trend |
| II. | WIFI wireless communication as an option | ||
| III. | Data cloud computing | ||
| IV. | Ethernet interface as standard supply for data transfer | ||
| V. | Remote operation and maintenance | ||
| VI. | Setting up a basic platform for interconnected tests | ||
| 2. | Real-time Diagnosis of instrument status | I. | Instrument’s communication, performance, and working status is monitored by real-time |
| 3. | Optional temperature stabilized, high sensitivity DLATGS detector module. | I. | Ensures high sensitivity, stability and good reproducibility of IR spectral analysis. Provides stable detector response |
| II. | Provides stable detector response |
| Sr.No | MainFTOS Suite Software | Function |
| 1. | All version Windows OS compatible | Windows XP, Windows 7, Windows 8, Windows 8.1 and Windows 10 (32 bits and 64 bits) |
| 2. | Modular platform design | Data collection, processing and analysis, report print out functions and different special software functions module options |
| 3. | FDA CFR 21 Part 11 Compliance | Access control, authority assignment, audit trial, digital signature |
| 4. | Automatic software updating | Getting the latest software version from the updating center |
| 5. | Real-time instrument status diagnosis | Real-time monitor of working status, specifications and communication status |
| 6. | Secondary development library | Providing C++ and C development sample code for different application development |
| 7. | Specialized IR libraries | Providing various specialized IR libraries |
| 8. | Touch control operation | Satisfying Window 10’s touch control function |
Fison FTIR Spectrometer range is designed for precise molecular and material analysis, offering spectral range from 7800 to 350 cm?¹ and high-resolution performance up to 2 cm?¹. These instruments ensure accurate spectral data with advanced interferometer technology and are suitable for research, quality control, and laboratory applications requiring reliable infrared analysis.