Single Beam UV-Visible Spectrophotometer FM-UVS-A200 is a compact unit comes with 4 nm of spectral bandwidth and 200 to 1000 nm of wavelength range. Features tungsten and deuterium lamp as light source with automatic switching and auto-setting of wavelength for fast calibration . Compact design with LCD display and 4 position sample compartments with 10 mm cuvette, it also has USB and printer interface.
Specifications
Wavelength range | 200 to 1000 nm |
Spectral bandwidth | 4 nm |
Optical system | Single beam |
Wavelength accuracy | ± 2 nm |
Wavelength repeatability | ≤ 1 nm |
Photometric range | -0.3 - 3.0 A, 0 to 200%T |
Photometric accuracy | ±0.003 A @ 0.5 A±0.006 A @ 1.0 A, ±0.5%T @ 100%T |
Light source | Tungsten lamp and deuterium lamp |
Stray light | ≤0.2%T @ 360nm |
Stability | ≤0.002 A/h @ 500 nm ( after preheat 2 hours) |
Display | LCD |
Standard cell holder | 4 positions |
Sample compartment | 10 mm path length cuvette |
Output | USB port and parallel port (printer) |
Power requirement | AC 110 / 220 V, 50 / 60 Hz |
Power | 120 W |
Dimensions (W × D × H) | 490 × 360 × 210 mm |
Weight | 14 kg |
Features
Applications
Accessories
Square glass cuvettes 10mm × 4 pc
Square quartz cuvettes 10 mm × 2 pc
Accessories Optional
Software
4 cell holders: 10 mm, 50 mm, 100 mm
Cylindrical cell holder
10 mm water jacketed holder: 1 cell and 4 cell
Micro cell holder
1 position Test tube holder: Φ8 to 22 mm, δ1.5 to 3 mm
8 position auto cell-changer
2 pc square glass cuvettes: 1 mm, 2 mm, 3 mm, 20 mm, 30 mm, 50 mm, 100 mm
2 pc square glass cuvettes: 1 mm, 2 mm, 3 mm, 20 mm, 30 mm, 50 mm, 100 mm
Fison Single Beam UV-Visible Spectrophotometers are designed for accurate measurement of light absorbance and transmittance. They provide a wide wavelength range, covering the UV and visible spectrum for diverse applications. These units include automatic wavelength adjustment, ensuring precise and consistent wavelength selection. Our Spectrophotometers retain data after power loss, ensuring secure results. They provide USB and printer connectivity, allowing easy data transfer and efficient reporting.